Uncertainties in Small-Angle Measurement Systems Used to Calibrate Angle Artifacts

Translation pending review.

比较自准直仪与相移 Fizeau 干涉测量,分析非理想反射面引入的小角度误差和被测量定义问题。

2004Journal of Research of the National Institute of Standards and Technology 109, 319–333full_textfull_text2026-11-17Jack A. Stone, Mohamed Amer, Bryon Faust, Jay Zimmerman

Original paper title retained. Summaries are editorial notes; obtain full text from the publisher or lawful open sources.

Findings

Finding

Translation pending review.

比较自准直仪与相移 Fizeau 干涉测量,分析非理想反射面引入的小角度误差和被测量定义问题。

Conditions

Translation pending review.

校准光学多面体等角度实物;比较系统采用不同的表面方向测量原理。

Source location

Translation pending review.

官方 PDF 第1页摘要、Introduction;DOI 经 NIST 109卷目录核对。

Application relevance

Translation pending review.

Translation pending review.

用于角度检测流程中明确反射面质量、有效光斑区域、表面方向定义与不确定度来源。

Limitations

Translation pending review.

Translation pending review.

不能仅凭两个仪器数值接近就认定系统无偏差;非平面反射面的平均方向需要明确定义。