SHANGHAI MICROCRE OPTICS-MECH TECH CO., LTDAbout MICROCRE400-812-7833
APPLICATION FILE / Composites

In-plane tension of continuous-fibre composites

A laminate needs 0° and 90° tensile modulus and strength to check the lay-up design.

Industry:CompositesInstruments:WDW-300M

The constraint

Grip ends concentrate stress and fail early, and extensometer knives can damage the surface.

Approach

Bond tabs, control grip pressure and measure strain in the middle section with a large-deformation extensometer at the standard rate.

  1. Bond gauges or fit the extensometer with the gauge length centred
  2. Load at 2 mm/min to failure and record curve features
  3. Compute modulus and strength per fibre direction and note failure modes

Outcome

0° modulus was within 4% of design while 90° strength fell short, prompting a lay-up change.

Related instruments

Application scenario files: industry, workpiece type and measurement goals come from routine engineering discussions. Customer names and specific data are anonymized; nothing here is a traceable commercial record or a commitment about any customer's production results.