Surface layer thickness and hardness on orthopaedic implants

Modified surface layers must be evaluated without damaging the functional face.

Industry:Medical devicesInstruments:HMAS-C1000SZA · MMAS-4 · ZMP-2000Z

The constraint

High-value, low-volume parts; indenting the functional face is not allowed.

Approach

Coupons processed with the parts; section and hardness on coupons, non-destructive thickness check on parts.

  1. Coupons processed in the same batch
  2. Coupon section thickness and interface hardness
  3. Non-destructive thickness spot check on parts

Outcome

A combined destructive-coupon plus non-destructive-part acceptance route avoided batch scrap risk.